Benutzer: Gast  Login
Dokumenttyp:
journal article 
Autor(en):
Schneider, FR; Mann, AB; Konorov, I; Delso, G; Paul, S; Ziegler, SI 
Titel:
A Positron Emission Tomograph based on LSO-APD modules with a sampling ADC read-out system for a students' advanced laboratory course. 
Abstract:
A one-day laboratory course on positron emission tomography (PET) for the education of physics students and PhD students in medical physics has been set up. In the course, the physical background and the principles of a PET scanner are introduced. Course attendees set the system in operation, calibrate it using a (22)Na point source and reconstruct different source geometries filled with (18)F. The PET scanner features an individual channel read-out of 96 lutetium oxyorthosilicate (LSO) scintill...    »
 
Zeitschriftentitel:
Z Med Phys 
Jahr:
2012 
Band / Volume:
22 
Heft / Issue:
Seitenangaben Beitrag:
143-9 
Sprache:
eng 
Print-ISSN:
0939-3889 
TUM Einrichtung:
Nuklearmedizinische Klinik und Poliklinik