Benutzer: Gast  Login
Dokumenttyp:
Journal Article; Article
Autor(en):
Jilek, C; Tzeis, S; Reents, T; Estner, HL; Fichtner, S; Ammar, S; Wu, J; Hessling, G; Deisenhofer, I; Kolb, C
Titel:
Safety of implantable pacemakers and cardioverter defibrillators in the magnetic field of a novel remote magnetic navigation system.
Abstract:
Electromagnetic interference with pacemaker and implantable cardioverter defibrillator (ICD) systems may cause temporary or permanent system malfunction of implanted devices. The aim of this study was to evaluate potential interference of a novel magnetic navigation system with implantable rhythm devices.A total of 121 devices (77 pacemakers, 44 ICDs) were exposed to an activated NIOBE II® Magnetic Navigation System (Stereotaxis, St. Louis, MO, USA) at the maximal magnetic field strength of 0.1...     »
Zeitschriftentitel:
J Cardiovasc Electrophysiol
Jahr:
2010
Band / Volume:
21
Heft / Issue:
10
Seitenangaben Beitrag:
1136-41
Sprache:
eng
Volltext / DOI:
doi:10.1111/j.1540-8167.2010.01806.x
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/20522155
Print-ISSN:
1045-3873
TUM Einrichtung:
I. Medizinische Klinik und Poliklinik (Kardiologie); Klinik für Kinderkardiologie und angeborene Herzfehler (Prof. Hess)
 BibTeX