Benutzer: Gast  Login
Dokumenttyp:
Journal Article; Multicenter Study; Validation Studies; Article
Autor(en):
Taborski, U; Braun, SL; Völler, H
Titel:
Analytical performance of the new coagulation monitoring system INRatio for the determination of INR compared with the coagulation monitor Coaguchek S and an established laboratory method.
Abstract:
An evaluation of the INRatio Prothrombin Time Monitoring system for determination of INR was done in two centers with a total of 5 healthy subjects and 77 subjects on oral anticoagulation. The INRatio and the Coaguchek S were compared with an established laboratory method. The correlation coefficient of the comparison with the laboratory was r=0.954 for INRatio and r=0.937 for Coaguchek S. The mean relative deviation from the lab method calculated according to Hill was 6.87% for INRatio, which i...     »
Zeitschriftentitel:
J Thromb Thrombolysis
Jahr:
2004
Band / Volume:
18
Heft / Issue:
2
Seitenangaben Beitrag:
103-7
Sprache:
eng
Volltext / DOI:
doi:10.1007/s11239-004-0206-8
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/15789176
Print-ISSN:
0929-5305
TUM Einrichtung:
Institut für Laboratoriumsmedizin (keine SAP-Zuordnung!)
 BibTeX