User: Guest  Login
Document type:
Journal Article; Research Support, Non-U.S. Gov't; Article
Author(s):
Mer?ep, Elena; Burton, Neal C; Claussen, Jing; Razansky, Daniel
Title:
Whole-body live mouse imaging by hybrid reflection-mode ultrasound and optoacoustic tomography.
Abstract:
We present a hybrid preclinical imaging scanner that optimally supports image acquisition in both reflection-mode ultrasonography and optoacoustic (OA) tomography modes. The system comprises a quasi-full-ring tomographic geometry capable of the simultaneous dual-mode imaging through entire cross sections of mice with in-plane spatial resolution in the range of 150 and 350 ?m in the respective OA and ultrasound (US) imaging modes with an imaging speed of up to 10 two-dimensional frames per second...     »
Journal title abbreviation:
Opt Lett
Year:
2015
Journal volume:
40
Journal issue:
20
Pages contribution:
4643-6
Language:
eng
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/26469584
Print-ISSN:
0146-9592
TUM Institution:
Lehrstuhl für Biologische Bildgebung - Zusammenarbeit mit dem Helmholtz-Zentrum München (Prof. Ntziachristos)
 BibTeX