Eichhorn, S; Steinhauser, E; Gradinger, R; Burgkart, R
Title:
New method for determining in vitro structure stiffness of ceramic acetabular liners under different impact conditions.
Abstract:
Increasing both patient mobility and prosthesis life span requires improvements in the range of motion and wear behavior of the liner. With the use of new composite alumina-zirconia ceramic materials, the same stability of the liner can be achieved at lower wall thickness than it is possible with alumina-only materials. The aim of this study was developing a method for determining the in vitro structure stiffness of ceramic acetabular liners against impact stresses. The first trials were perform...     »
Journal title abbreviation:
Med Eng Phys
Year:
2012
Journal volume:
34
Journal issue:
4
Pages contribution:
512-5
Language:
eng
Fulltext / DOI:
doi:10.1016/j.medengphy.2011.11.020
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/22197061
Print-ISSN:
1350-4533
TUM Institution:
Klinik für Herz- und Gefäßchirurgie (Prof. Lange); Klinik und Poliklinik für Orthopädie und Sportorthopädie