Benutzer: Gast  Login
Dokumenttyp:
journal article 
Autor(en):
Gahlert, M; Burtscher, D; Grunert, I; Kniha, H; Steinhauser, E 
Titel:
Failure analysis of fractured dental zirconia implants. 
Abstract:
The purpose of the present study was the macroscopic and microscopic failure analysis of fractured zirconia dental implants.Thirteen fractured one-piece zirconia implants (Z-Look3) out of 170 inserted implants with an average in situ period of 36.75±5.34 months (range from 20 to 56 months, median 38 months) were prepared for macroscopic and microscopic (scanning electron microscopy [SEM]) failure analysis. These 170 implants were inserted in 79 patients. The patient histories were compared with...    »
 
Zeitschriftentitel:
Clin Oral Implants Res 
Jahr:
2012 
Band / Volume:
23 
Heft / Issue:
Seitenangaben Beitrag:
287-93 
Sprache:
eng 
Print-ISSN:
0905-7161 
TUM Einrichtung:
dische Klinik und Poliklinik