Impairment of pericardial leaflet structure from balloon-expanded valved stents

Autor(en):
de Buhr, W., Pfeifer, S., Slotta-Huspenina, J., Wintermantel, E., Lutter, G., Goetz, W. A.
Abstract:
Objective: Malpositioning is one of the major problems in transcatheter aortic valve implantation. To evaluate the influence of mechanical balloon inflation on aortic valve stent positioning, the expansion process and the impact on the valve leaflet's structure were investigated. Methods: Custom-made stents were laser cut from a 22-mm diameter stainless steel tube and mounted with a glutaraldehyde-treated bovine pericardial valve. The valved stents were crimped onto a standard balloon catheter a...     »
Stichworte:
aortic bioprosthetic valves calcification implantation substitutes replacement
Kongresstitel:
ISI Document Delivery No.: 943HE Times Cited: 3 Cited Reference Count: 17 de Buhr, Wiebke Pfeifer, Stefan Slotta-Huspenina, Julia Wintermantel, Erich Lutter, Georg Goetz, Wolfgang A. Mosby-elsevier New york
Zeitschriftentitel:
Journal of Thoracic and Cardiovascular Surgery
Jahr:
2012
Band / Volume:
143
Heft / Issue:
6
Seitenangaben Beitrag:
1417-1421
Nachgewiesen in:
Scopus
Reviewed:
ja
Volltext / DOI:
doi:10.1016/j.jtcvs.2011.11.001