User: Guest  Login
Document type:
journal article 
Author(s):
Khoramnia, R; Lohmann, CP; Wuellner, C; Kobuch, KA; Donitzky, C; Winkler von Mohrenfels, C 
Title:
Effect of 3 excimer laser ablation frequencies (200 Hz, 500 Hz, 1000 Hz) on the cornea using a 1000 Hz scanning-spot excimer laser. 
Abstract:
To evaluate the possible side effects on human and porcine cadaver eyes induced by excimer laser ablation with 3 ablation frequencies.Augenklinik, Klinikum rechts der Isar, Technische Universitaet Muenchen, Munich, Germany.Central epithelial abrasion was performed on porcine and human donor eyes. Next, the porcine eyes were photoablated (-9.00 diopters) at 1 of 3 frequencies (200 Hz, 500 Hz, or 1000 Hz) using a prototype 1000 Hz excimer laser. The human eyes were ablated at 1000 Hz. The surface...    »
 
Journal title abbreviation:
J Cataract Refract Surg 
Year:
2010 
Journal volume:
36 
Journal issue:
Pages contribution:
1385-91 
Language:
eng 
Print-ISSN:
0886-3350 
TUM Institution:
Augenklinik und Poliklinik