First Robotic SPECT for Minimally Invasive Sentinel Lymph Node Mapping
Abstract:
In this paper we present the usage of a dropin gamma probe for intra-operative Single-Photon Emission Computed Tomography (SPECT) imaging in the scope of minimally invasive robot-assisted interventions. The probe is designed to be inserted and reside inside the abdominal cavity during the intervention. It is grasped during the procedure using a robotic laparoscopic gripper enabling full six degrees of freedom handling by the surgeon. We demonstrate the first deployment of the tracked probe for i...     »
Keywords:
Lymph nodes; Probes; Robots; Single photon emission computed tomography; Surgery; Tracking; Endoscopy/Laparoscopy; Image-guided treatment; SPECT; Surgical guidance/navigation
Journal title:
Medical Imaging, IEEE Transactions on
Year:
2015
Journal volume:
Pp
Journal issue:
99
Pages contribution:
1--1
Print-ISSN:
0278-0062
 BibTeX