Original title:
Isotropic 3D fast spin-echo imaging versus standard 2D imaging at 3.0 T of the knee—image quality and diagnostic performance
Translated title:
Isotrope 3D fast spin-echo Bildgebung gegenüber standardisierter 2D Bildgebung des Knies mit 3.0 T — Bildqualität und Diagnostische Leistung
Author:
Ristow, Oliver
Year:
2011
Document type:
Dissertation
Faculty/School:
Fakultät für Medizin
Advisor:
Link, Thomas M. (Prof. Dr.)
Language:
en
Subject group:
MED Medizin
Keywords:
Knee-studies, Comparative-MRI, Sequences-3D fast spin-echo imaging
Abstract:
The objective of this study was to compare a newly developed fat-saturated intermediate-weighted 3D fast spin-echo sequence with standard 2D IM-w FSE sequences regarding image quality and diagnostic performance in assessing abnormal findings of the knee. MR imaging was performed at 3.0 T in 50 patients. Images were assessed independently by three radiologists. Image quality was rated significantly higher for the 2D versus the 3D FSE sequences. Sensitivity for cartilage lesions was slightly highe...     »
Translated abstract:
Das Ziel dieser Studie war es, eine neu entwickelte fettgesättigt-intermediär gewichtete 3D fast spin-echo Sequenz mit einer in der klinischen Routine verwendeten 2D IM-w FSE Sequenz hinsichtlich ihrer Effizienz im Erkennen pathologischer Veränderungen des Knies zu vergleichen. Von 50 Patienten wurden MRT Bilder mit 3.0T aufgenommen und durch 3 Radiologen befundet. Im Ergebnis war die Bildqualität für die 2D Sequenzen signifikant besser. In der Erkennung von Knorpelverletzungen war die 3D Sequen...     »