User: Guest  Login
Original title:
Silicon photonics sensors of ultrasound for optoacoustic imaging 
Translated title:
Silizium-Photonik Ultraschallsensoren für die optoakustische Bildgebung 
Year:
2021 
Document type:
Dissertation 
Institution:
Fakultät für Elektrotechnik und Informationstechnik 
Advisor:
Ntziachristos, Vasilis (Prof. Dr.) 
Referee:
Ntziachristos, Vasilis (Prof. Dr.); Wolfrum, Bernhard (Prof. Dr.) 
Language:
en 
Subject group:
FEI Feinwerktechnik, Medizintechnik, Technische Optik, Reprographietechnik; MED Medizin 
Keywords:
Integrated optics, optoacoustics, photoacoustics, imaging, microscopy, sensors, ultrasound 
TUM classification:
MED 385d 
Abstract:
The resolution of optoacoustic imaging is determined by the aperture and the bandwidth of the ultrasound sensor. Silicon-photonics technology is used to develop a sensor with an aperture of 220 × 500 square nanometers and a bandwidth of 280 MHz. The sensor improves image resolution over state-of-the-art piezoelectric sensors and enables near-field optoacoustic imaging with λ/50 super-resolution. The near-field images visualize absorption contrast with optical resolutions of 650 nanometers. 
Translated abstract:
Die Auflösung optoakustischer Bildgebung wird durch die Apertur und Bandbreite des Ultraschallsensors bestimmt. Silizium-Photonik Technologie ist hier zur Entwicklung eines Sensors mit einer Apertur von 220 x 500 nm² und einer Bandbreite von 280 MHz benutzt. Der Sensor verbessert die Bildauflösung jenseits modernster piezoelektrischer Sensoren und erlaubt Nahfeld optoakustische Bildgebung mit λ/50 Hochauflösung, die Absorptionskontrast mit Auflösungen von 650 nm aufzeigt. 
Oral examination:
10.05.2021 
File size:
10453859 bytes 
Pages:
151 
Last change:
29.07.2021