Titel:

Precision measurement of cardiac structure and function in cardiovascular magnetic resonance using machine learning.

Dokumenttyp:
Journal Article
Autor(en):
Davies, Rhodri H; Augusto, João B; Bhuva, Anish; Xue, Hui; Treibel, Thomas A; Ye, Yang; Hughes, Rebecca K; Bai, Wenjia; Lau, Clement; Shiwani, Hunain; Fontana, Marianna; Kozor, Rebecca; Herrey, Anna; Lopes, Luis R; Maestrini, Viviana; Rosmini, Stefania; Petersen, Steffen E; Kellman, Peter; Rueckert, Daniel; Greenwood, John P; Captur, Gabriella; Manisty, Charlotte; Schelbert, Erik; Moon, James C
Abstract:
BACKGROUND: Measurement of cardiac structure and function from images (e.g. volumes, mass and derived parameters such as left ventricular (LV) ejection fraction [LVEF]) guides care for millions. This is best assessed using cardiovascular magnetic resonance (CMR), but image analysis is currently performed by individual clinicians, which introduces error. We sought to develop a machine learning algorithm for volumetric analysis of CMR images with demonstrably better precision than human analysis. METHODS: A fully automated machine learning algorithm was trained on 1923 scans (10 scanner models, 13 institutions, 9 clinical conditions, 60,000 contours) and used to segment the LV blood volume and myocardium. Performance was quantified by measuring precision on an independent multi-site validation dataset with multiple pathologies with n = 109 patients, scanned twice. This dataset was augmented with a further 1277 patients scanned as part of routine clinical care to allow qualitative assessment of generalization ability by identifying mis-segmentations. Machine learning algorithm ('machine') performance was compared to three clinicians ('human') and a commercial tool (cvi42, Circle Cardiovascular Imaging). FINDINGS: Machine analysis was quicker (20 s per patient) than human (13 min). Overall machine mis-segmentation rate was 1 in 479 images for the combined dataset, occurring mostly in rare pathologies not encountered in training. Without correcting these mis-segmentations, machine analysis had superior precision to three clinicians (e.g. scan-rescan coefficients of variation of human vs machine: LVEF 6.0% vs 4.2%, LV mass 4.8% vs. 3.6%; both P < 0.05), translating to a 46% reduction in required trial sample size using an LVEF endpoint. CONCLUSION: We present a fully automated algorithm for measuring LV structure and global systolic function that betters human performance for speed and precision.
Zeitschriftentitel:
J Cardiovasc Magn Reson
Jahr:
2022
Band / Volume:
24
Heft / Issue:
1
Volltext / DOI:
doi:10.1186/s12968-022-00846-4
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/35272664
Print-ISSN:
1097-6647
TUM Einrichtung:
Institut für KI und Informatik in der Medizin
 BibTeX