Benutzer: Gast  Login
Dokumenttyp:
Journal Article
Autor(en):
Ehn, Sebastian; Sellerer, Thorsten; Muenzel, Daniela; Fingerle, Alexander A; Kopp, Felix; Duda, Manuela; Mei, Kai; Renger, Bernhard; Herzen, Julia; Dangelmaier, Julia; Schwaiger, Benedikt J; Sauter, Andreas; Riederer, Isabelle; Renz, Martin; Braren, Rickmer; Rummeny, Ernst J; Pfeiffer, Franz; Noël, Peter B
Titel:
Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system.
Abstract:
The performance of a recently introduced spectral computed tomography system based on a dual-layer detector has been investigated. A semi-anthropomorphic abdomen phantom for CT performance evaluation was imaged on the dual-layer spectral CT at different radiation exposure levels (CTDIvol of 10 mGy, 20 mGy and 30 mGy). The phantom was equipped with specific low-contrast and tissue-equivalent inserts including water-, adipose-, muscle-, liver-, bone-like materials and a variation in iodine concent...     »
Zeitschriftentitel:
J Appl Clin Med Phys
Jahr:
2018
Band / Volume:
19
Heft / Issue:
1
Seitenangaben Beitrag:
204-217
Volltext / DOI:
doi:10.1002/acm2.12243
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/29266724
TUM Einrichtung:
Institut für Diagnostische und Interventionelle Radiologie
 BibTeX