Benutzer: Gast  Login
Dokumenttyp:
Evaluation Studies; Journal Article; Validation Studies
Autor(en):
Bader, R; Steinhauser, E; Scholz, R; Simnacher, M; Mittelmeier, W
Titel:
[Experimental analysis of neutral, asymmetric and constraint liners for total hip replacement: investigation of range of motion and protection against joint instability]
Abstract:
AIM: Recurrent dislocation after total hip replacement is a severe complication, which requires specific treatment and implants. The purpose of the present study was to compare a constraint liner with an elevated rim and standard liner regarding their range of motion and dislocation stability. METHOD: With a test device, range of motion until impingement (ROM (Imp)) and dislocation (ROM (Lux)) were experimentally analyzed using the above-mentioned insert types of a commercial total hip system. O...     »
Zeitschriftentitel:
Z Orthop Ihre Grenzgeb
Jahr:
2004
Band / Volume:
142
Heft / Issue:
5
Seitenangaben Beitrag:
577-85
Sprache:
de
Volltext / DOI:
doi:10.1055/s-2004-832318
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/15472768
Print-ISSN:
0044-3220
TUM Einrichtung:
Klinik und Poliklinik für Orthopädie und Sportorthopädie
 BibTeX