User: Guest  Login
Original title:
Optoacoustic handheld imaging for clinical screening and intervention
Translated title:
Handgeführte optoakustische Bildgebung für klinische Früherkennung und Intervention
Author:
Dima, Alexander
Year:
2016
Document type:
Dissertation
Faculty/School:
Fakultät für Elektrotechnik und Informationstechnik
Advisor:
Ntziachristos, Vasilis (Prof., Ph.D.)
Referee:
Ntziachristos, Vasilis (Prof., Ph.D.); Diepold, Klaus (Prof. Dr.)
Language:
en
Subject group:
FEI Feinwerktechnik, Medizintechnik, Technische Optik, Reprographietechnik
Keywords:
Bildgebung, Medizin, Optisch
TUM classification:
MED 385d
Abstract:
A handheld probe for non-invasive optoacoustic imaging has been developed for clinical diagnostic imaging. The probe enables visualization of optical absorption contrast at tissue depths up to 30 mm below the skin surface with resolutions of up to 150µm. Initial results from healthy human volunteers reveal optical features of the thyroid, the cardio-vascular system and of muscles at resolutions previously not available. A second configuration and method, designed for intraoperative assessment of...     »
Translated abstract:
Eine handgeführte Geometrie für die nicht-invasive optoakustische Bildgebung in der klinischen Diagnostik wurde entwickelt. Die Sonde ermöglicht die Visualisierung optischer Absorption (primärer Kontrast) in Gewebe mit Eindringtiefen unter der Haut von bis zu 30 mm und einer Auflösung von bis zu 150 µm. Erste Ergebnisse am gesunden Menschen enthüllen optische Eigenschaften der Schilddrüse, des kardio-vaskulären System und der Muskeln mit bisher unerreichter Auflösung. Eine zweite Konfiguration u...     »
WWW:
https://mediatum.ub.tum.de/?id=1251669
Date of submission:
22.06.2016
Oral examination:
29.01.2016
File size:
8309412 bytes
Pages:
146
Urn (citeable URL):
https://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:91-diss-20160129-1251669-1-8
Last change:
18.03.2016
 BibTeX