Benutzer: Gast  Login
Dokumenttyp:
Journal Article
Autor(en):
Kreuser, Laura; Laugwitz, Karl-Ludwig; Tiemann, Klaus; Lewalter, Thorsten; Jilek, Clemens
Titel:
Preparation of PCI Balloons: What Is the Best Method to Avoid Air in the Balloon? A Comparison of Different Methods of Connecting PCI Balloons and the Inflation Syringe while Removing Air from the Balloon.
Abstract:
As the techniques to connect percutaneous coronary intervention (PCI) balloons and the inflation syringe vary in the instructions for use and in practice, we measured the amount of air in PCI balloons after testing three connection methods to an inflation syringe. Following the preparation using one of the three methods, 114 balloons and stent balloons were tested four times. Method 1 connected the syringe and the balloon catheter directly after purging and filling the lumen, while method 3 omit...     »
Zeitschriftentitel:
J Clin Med
Jahr:
2020
Band / Volume:
9
Heft / Issue:
1
Volltext / DOI:
doi:10.3390/jcm9010172