Abstract:
Geometric tomography and conventional algebraic tomography algorithms are used to reconstruct cross-sections of an InAs nanowire from a tilt series of experimental annular dark-field images. Both algorithms are also applied to a test object to assess what factors affect the reconstruction quality. When using the present algorithms, geometric tomography is faster, but artifacts in the reconstruction may be difficult to recognize.