Kolb, C; Zrenner, B; Schreieck, J; Schmitt, C
Titel:
Strategies to avoid inappropriate therapies due to ventricular double detection in biventricular pacing implantable cardioverter/defibrillators.
Abstract:
Implantation of biventricular pacing implantable cardioverter/defibrillators (ICD) is a new therapeutic approach for the treatment of chronic heart failure and the prevention of sudden cardiac death. Due to a common left and right ventricular sensing channel in first generation biventricular pacing ICD, ventricular double sensing and subsequent delivery of inappropriate therapies occurs in some patients. A case of inappropriate anti-tachycardia pacing due to ventricular double detection is descr...     »
Zeitschriftentitel:
Eur J Heart Fail
Jahr:
2003
Band / Volume:
5
Heft / Issue:
4
Seitenangaben Beitrag:
523-6
Sprache:
eng
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/12921814
Print-ISSN:
1388-9842
TUM Einrichtung:
I. Medizinische Klinik und Poliklinik (Kardiologie)
 BibTeX