Benutzer: Gast  Login
Dokumenttyp:
Article
Autor(en):
Mandl, S; Sader, R; Thorwarth, G; Krause, D; Zeilhofer, HF; Horch, HH; Rauschenbach, B
Titel:
Biocompatibility of titanium based implants treated with plasma immersion ion implantation
Abstract:
In this work, the biocompatibility of titanium before and after oxygen PIII is investigated using a rat animal model. Pure titanium (grade 2) and pre-anodized titanium were implanted with oxygen at elevated temperatures between 200 and 550 degreesC and subsequently analyzed for oxygen content and phase composition. No deterioration of the tensile strength and the yield strength was detected after the implantation. The mechanical stability of the osseointegration was determined with a pull-out te...     »
Zeitschriftentitel:
Nucl Instrum Methods Phys Res B
Jahr:
2003
Band / Volume:
206
Seitenangaben Beitrag:
517-521
Sprache:
eng
Print-ISSN:
0168-583X
TUM Einrichtung:
Klinik und Poliklinik für Mund-, Kiefer- und Gesichtschirurgie
 BibTeX