Benutzer: Gast  Login
Dokumenttyp:
Journal Article
Autor(en):
Gahlert, M; Burtscher, D; Grunert, I; Kniha, H; Steinhauser, E
Titel:
Failure analysis of fractured dental zirconia implants.
Abstract:
The purpose of the present study was the macroscopic and microscopic failure analysis of fractured zirconia dental implants.Thirteen fractured one-piece zirconia implants (Z-Look3) out of 170 inserted implants with an average in situ period of 36.75±5.34 months (range from 20 to 56 months, median 38 months) were prepared for macroscopic and microscopic (scanning electron microscopy [SEM]) failure analysis. These 170 implants were inserted in 79 patients. The patient histories were compared with...     »
Zeitschriftentitel:
Clin Oral Implants Res
Jahr:
2012
Band / Volume:
23
Heft / Issue:
3
Seitenangaben Beitrag:
287-93
Sprache:
eng
Volltext / DOI:
doi:10.1111/j.1600-0501.2011.02206.x
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/21545535
Print-ISSN:
0905-7161
TUM Einrichtung:
Klinik und Poliklinik für Orthopädie und Sportorthopädie
 BibTeX